Abstract

An automated and miniaturized chemiluminescence method for screening of fluoroquinolones in milk samples was proposed. The method was based on magnetic dispersive micro-solid phase extraction of analytes followed by the chemiluminescence determination of the total fluoroquinolones content using a multi-pumping flow system. In the developed method, Zr-Fe-C magnetic nanoparticles were used as an efficient sorbent for separation of fluoroquinolones from sample matrix. The chemiluminescence intensity obtained for Ce4+-SO32−-Tb3+ chemiluminescence system was greatly increased in the presence of the analytes. Under the optimal conditions, the detector response for fluoroquinolones was linear in the concentration ranges from 5·10−9 to 1·10−6 mol L−1 with respect to fleroxacin. The limit of detection, calculated from a blank test based on 3σ, was 2·10−9 mol L−1 with respect to fleroxacin. The presented method demonstrated to be a good tool for available and cost-effective point-of-need screening fluoroquinolones in milk samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.