Abstract

Fat-suppression techniques are used extensively in routine proton nuclear magnetic resonance imaging to produce images free from chemical shift artifacts and dynamic range problems. A hybrid fat-suppression sequence is studied which combines the principle of short time inversion recovery with chemical shift selective imaging. The aim of this study is to provide a theoretical understanding of the role of the sequence parameters, as well as to compare this hybrid sequence with its most closely related conventional fat-suppression techniques, namely selective presaturation and short time inversion recovery (STIR) imaging. The hybrid technique is shown to be robust in normal use, and more tolerant than the conventional methods to mis-settings of parameters such as inversion time, as well as tip angle and frequency bandwidth of the fat selective pulse.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.