A three-dimensional verification method using dye gel dosimeters is studied for the external exposure in boron neutron capture therapy (BNCT). The characterization experiments of the dye gel dosimeters were performed for the dose response to gamma rays. As the dye gel dosimeters, leucomalite-green, leucocrystal-violet and genipin gels, which have different coloration characteristics, were selected. The results for the dose response to gamma rays showed linearity up to 100 Gy. However, it was found that the correction for auto-oxidation and dose rate dependency were necessary.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.