Twiddler syndrome (TS) is a complication of cardiac implantable electronic device (CIED) implantation, caused by the deliberate or unconscious manipulation of the device by the patient himself, which results in dislocation of the leads by retraction towards the subcutaneous pocket. This report describes two clinical cases that occurred in our centre, for which two different solutions were successfully implemented. In the first case, a complete removal of the stimulation system was performed, and a leadless pacemaker (PM; Medtronic Micra VR) was implanted. In the second case, the patient underwent a revision procedure. The PM was disconnected, and the electrodes were debrided, a submuscular pocket for the PM was created, and at the end of the procedure, the PM was anchored to the pectoralis major. Twiddler syndrome is a not so rare and serious complication of CIED implantation, leading to device malfunctioning and higher risk of infection of the pocket due to multiple re-interventions. In these two cases, different surgical solutions were performed, both resulting to be effective to solve the effects of TS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.