Abstract

A case of spontaneous fracture of a catheter of a totally implantable venous access port is described. A totally implantable venous access port was placed in a 62-year-old male patient with short bowel syndrome for home parenteral nutrition. Thirteen months after implantation it was noted on chest X-ray that the catheter had fractured and the distal fragment embolized to the right atrium. The embolized catheter fragment was retrieved viaa percutaneous femoral vein approach using a basket catheter. The catheter inserted using the percutaneous subclavian puncture technique has been broken due to the long-standing compression at the narrow space between the clavicle and first rib. Catheter fracture is a relatively rare, but avoidable complication by introducing the catheter lateral to the midclavicular point.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.