Abstract

A capacitive sensor for 2,4-dichloro phenoxy acetic acid(2,4-D) determination in drinking water has been developed using molecularly imprinted polypyrrole on pencil graphite electrode (PGE). Molecular imprinted polymer (MIP) coated PGE was prepared by electropolymerization of pyrrole via chronopotentiometry in the presence of 2,4-D as the template molecule. The prepared electrodes were characterized by field emission gun-scanning electron microscopy, cyclic voltammetry, and electrochemical impedance spectroscopy (EIS). The capacitance change of MIP electrode was measured in the presence of 2,4-D using EIS. The developed capacitive sensor exhibited a linear range 0.06–1.25 µg l−1 2,4-D with limit of detection of 0.02 µg l−1 and good selectivity towards 2,4-D in water with recovery from 92 to 110%. The results suggest the viable applicability of the MIP/PGE based sensor for the determination of the 2,4-D in water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.