Abstract

Abstract A cantilever array sensor platform was developed based on the optical lever method, and the cantilever array chip was also fabricated to introduce its applications on biochemical detection. Optical fibers coupled to lasers were used as the scanning light source. This sensor system had good stability, and the noise of the detection signal was about 2 nm. Meanwhile, the cantilevers of the fabricated chip had good straightness and were consistent to temperature response, and the deviations of the response sensitivity for cantilevers' temperature change were no more than 5.0%. This sensing system was used to detect Hg 2+ in aqueous solution within a concentration range of 1–200 ng mL −1 . The deflections of one array chip were close in the same concentration, and the average deviation was less than 15%. The samples with the concentrations of 1.0 and 0.2 ng mL −1 Hg 2+ were detected separately by the fabricated chip and a foreign commercial chip on the sensor platform. The result showed that it was difficult for the fabricated chip to achieve a higher detection sensitivity and necessary to improve the cantilever array production process.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.