Abstract
In this paper we report a calorimetrie sensing system for the detection of clinically relevant biomolecules. The calorimetrie biosensor consists of a thin (∼20 μm) micromachined Y-cut quartz crystal resonator (QCR) placed in close proximity to a fluidic chamber packed with an immobilized enzyme in alginate beads. This configuration enables a detection mechanism physically separated from the solution of interest and is thereby free from fouling. The performance of this biosensor is demonstrated by detection of urea, creatinine and hydrogen peroxide in phosphate buffer via a flow injection analysis (FIA) technique. This work demonstrates a sensing modality in which the sensor itself is not fouled or contaminated by the solution of interest. Such a system enables reuse and reliability for long term sampling measurements. Further, the immobilized enzyme is easily discarded and replaced with a new reaction column. We envision a biosensing system which can perform rapid measurements to detect biomarkers such as glucose, creatinine, urea and lactate in urine and blood continuously over extended periods of time.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.