Abstract

In this paper a calibration study of the Dupont 910 DSC module is presented. According to our view of the cell construction, this type of calorimeter combines the advantages and disadvantages of a DTA and DSC cell. It was found that the calibration constant E is independent of temperature (as it should be for any DSC cell) but is dependent on the heating rate and the mass or the total absolute heat exchange by the sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.