Abstract
The self-sustaining atomic magnetometer has the advantage of high sensitivity and long spin coherence lifetime, based on which we demonstrate a novel method to calibrate the magnetic coil constants precisely. Via non-destructive phase measurement and coherent optical pumping, the spin polarization of rubidium atoms is regenerated coherently and the Larmor precession signal is oscillating continuously. In this stable state, the calibrating capability is achieved by applying current to coils and scan the magnetic components along x- and y-direction. The magnetic field magnitude is obtained from precession frequency and the coil constants can be derived from the fitting equation directly. The constants of coils in the experiment are 246.010 ± 0.034 nT/mA and 197.452 ± 0.025 nT/mA in the x- and y-directions, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.