Abstract
This paper presents a method for extrapolating the total body water (TBW) resistance Rt50 from the resistance measured at 50 kHz (R50). A DXA examination and impedance measurements were carried out in a 1st group of 57 healthy volunteers with a Xitron 4200 multifrequency impedancemeter, in order to determine their values of Rt50 by comparison with resistances extrapolated at an infinite frequency by the Xitron (R∞). TBW volumes were calculated using our modified BIS method (Jaffrin et al 2006 Med. Biol. Eng. Comput. 44 873–82) from R∞, Rt50 and from the fat-free mass measured by DXA, assuming a hydration rate of 73.2%. The same protocol and calculations were also carried out on a 2nd group of 21 subjects for independent validation. Data of the 1st group showed that values of Rt50, not significantly different from those of R∞, could be obtained by dividing R50 by 1.231 in men and by 1.224 in women. Applying this method to the 2nd group yielded also values of Rt50 not significantly different from R∞. TBW volumes Vt50 obtained from Rt50 were not significantly different from those of our modified BIS method Vtn, or from TBW volumes obtained from DXA in both groups. A comparison with three BIA methods of TBW determination showed that our new method gave results in better agreement with TBW from DXA and from our modified BIS method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.