Abstract

A new amperometric biosensor based on urate oxidase–peroxidase coupled enzyme system for the specific and selective determination of uric acid in urine was developed. Commercially available urate oxidase and peroxidase were immobilized with gelatin by using glutaraldehyde and fixed on a pretreated teflon membrane. The method is based on generation of H 2O 2 from urine uric acid by urate oxidase and its consuming by peroxidase and then measurement of the decreasing of dissolved oxygen concentration by the biosensor. The biosensor response depends linearly on uric acid concentration between 0.1 and 0.5 μM. In the optimization studies of the biosensor, phosphate buffer (pH 7.5; 50 mM) and 35 °C were obtained as the optimum working conditions. In addition, the most suitable enzyme activities were found as 64.9×10 −3 U cm −2 for urate oxidase and 512.7 U cm −2 for peroxidase. And also some characteristic studies of the biosensor such as reproducibility, substrate specificity and storage stability were carried out.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.