Abstract

Schuhli locking nuts can be used in poor quality cortical bone to enhance fixation stability as an alternative to cement augmented screws. This study compared the fixation strength and stability of plate constructs using Schuhli locking nuts with standard screws and cement augmented screws for fixation of simulated humeral shaft fractures in a test model with osteoporosis. The constructs were tested in axial compression, 4-point bending, and torsion to determine fixation stability. The humeri were cycled in torsion (4.5 Nm) for 1000 cycles to simulate upper extremity use during the early postoperative period and retested for stability. The Schuhli locking nuts and cement augmented screws had significantly greater fixation stability than the standard screws before (range, 6-14 times greater) and after cycling in torsional loading (range, 3-3.6 times greater). Although cement augmented screws and Schuhli augmentation showed increased fixation stability compared with the standard screws in axial and 4-point bending before cycling (range, 1.3-1.4 times greater), this was not significant. Compared with Schuhli fixation, cement augmented screws showed no significant difference in fixation stability in all loading modes before and after cycling. Schuhli locking nuts offer the stability of cement augmentation while avoiding its potential adverse effects on fracture healing with extravasation and thermal necrosis.