Abstract

Seven-hole 3.5 mm broad and 5-hole 4.5 mm narrow dynamic compression plates were applied to paired canine cadaveric tibias in a stable fracture model. Paired tibias were tested to acute failure in rotation and four-point bending, and to fatigue failure in four-point bending. Resistance to screw pullout was measured for three 3.5 mm cortical screws and two 4.5 mm cortical screws inserted in the configurations of the bone plates. All plate-bone systems failed by fracture of the bone through a screw hole. The 3.5 mm plate-bone system was stronger in acute failure in rotation and in four-point bending. There was no difference in stiffness, and no difference in the number of cycles to failure in fatigue testing. Three 3.5 mm screws had greater resistance to pullout than two 4.5 mm screws. Results indicate that the 7-hole 3.5 mm broad dynamic compression plate has a biomechanical advantage over the 5-hole 4.5 mm narrow dynamic compression plate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.