Abstract

3.5 locking compression plate (LCP) fixation was compared to 3.5 limited contact dynamic compression plate (LC-DCP) fixation in a canine cadaveric, distal humeral metaphyseal gap model. Thirty paired humeri from adult, large breed dogs were separated into equal groups based on testing: static compression, cyclic compression, and cyclic torsion. Humeral constructs stabilised with LCP were significantly stiffer than those plated with LC-DCP when loaded in static axial compression (P = 0.0004). When cyclically loaded in axial compression, the LCP constructs were significantly less stiff than the LC-DCP constructs (P = 0.0029). Constructs plated with LCP were significantly less resistant to torsion over 500 cycles than those plated with LC-DCP (P<0.0001). The increased stiffness of LCP constructs in monotonic loading compared to constructs stabilised with non-locking plates may be attributed to the stability afforded by the plate-screw interface of locking plates. The LCP constructs demonstrated less stiffness in dynamic testing in this model, likely due to plate-bone offset secondary to non-anatomic contouring and occasional incomplete seating of the locking screws when using the torque-limiting screw driver. Resolution of these aspects of LCP application may help improve the stiffness of fixation in fractures modeled by the experimental set-up of this investigation.