Abstract

Fecal radiobioassay is an essential and sensitive tool to estimate the internal intake of actinides after a radiological incident. A new fecal analysis method, based on lithium metaborate fusion of fecal ash for complete sample dissolution followed by sequential column chromatography separation of actinides, has been developed for the determination of low-level Am and Cm in a large size sample. Spiked synthetic fecal samples were analyzed to evaluate method performance against the acceptance criteria for radiobioassay as defined by ANSI N13.30; both satisfactory accuracy and repeatability were achieved. This method is a promising candidate for reliable dose assessment of low level actinide exposure to meet the regulatory requirements of routine radiobioassay for nuclear workers and the public.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.