Abstract
A bilateral comparative study between two metrology institutes was conducted for the determination of ethanol in bioethanol fuel by isotope dilution mass spectrometry. Measurement procedures were independently developed by the Health Sciences Authority (Singapore) and the National Institute of Metrology (People’s Republic of China). The methods used included headspace gas chromatography–high-resolution mass spectrometry, headspace gas chromatography–mass spectrometry (GC-MS) and direct injection to GC-MS. An investigation of all potential sources of uncertainty, including a component of uncertainty due to the effect of using different instruments such as gas chromatography–flame ionization detector method, commonly adopted for the analysis of organic volatiles, was carried out. The results of the analyses are consistent within the associated measurement uncertainty of each laboratory, reported at approximately 95% confidence level.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.