Abstract

A 16-section BGO scintillation detector having a total volume of 8.54 ℓ and an associated 40 MHz flash-ADC-based data taking system were developed. This detector system is intended for measurements of neutron capture cross-sections of radioactive nuclei by the time-of-flight method. The detector response function and efficiency, gamma-ray and neutron energy resolution, and backgrounds were evaluated experimentally and by calculation. Neutron capture cross-section measurements with 10 B , 197 Au and 237 Np samples were carried out in the energy range of 1– 1000 eV to demonstrate the performance of the system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.