Abstract

The background of CR-39 detectors varies from batch to batch, from foil to foil in the same batch, from one side to another of the same foil and within the same foil surface. In spite of many efforts made in the past, little success has been achieved in producing detectors with a consistently low background. For this reason, new investigations have been undertaken with the specific scope in mind to improve the background of these detectors. The key strategy of these studies was to investigate whether the background tracks originate from the detector surface or from detector bulk, if not from both. From these systematic investigations it was possible to demonstrate that most of the background is due to defects present on the detector surface. For this reason, the simplest way to reduce the background is to carry out a heavy pre-etching prior to the use of the CR-39 detector. This procedure has been successfully exploited not only to increase the signal-to-noise ratio of CR-39-based radon dosimeters but also to carry out different radon measurements with the same CR-39 foil.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.