Abstract
Systems that automate extraction of nucleic acid from cells or viruses in complex clinical matrices have tremendous value even in the absence of an integrated downstream detector. We describe our bench-top automated workstation that integrates our previously-reported extraction method – TruTip – with our newly-developed mechanical lysis method. This is the first report of this method for homogenizing viscous and heterogeneous samples and lysing difficult-to-disrupt cells using "MagVor": a rotating magnet that rotates a miniature stir disk amidst glass beads confined inside of a disposable tube. Using this system, we demonstrate automated nucleic acid extraction from methicillin-resistant Staphylococcus aureus (MRSA) in nasopharyngeal aspirate (NPA), influenza A in nasopharyngeal swabs (NPS), human genomic DNA from whole blood, and Mycobacterium tuberculosis in NPA. The automated workstation yields nucleic acid with comparable extraction efficiency to manual protocols, which include commercially-available Qiagen spin column kits, across each of these sample types. This work expands the scope of applications beyond previous reports of TruTip to include difficult-to-disrupt cell types and automates the process, including a method for removal of organics, inside a compact bench-top workstation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.