Abstract
AbstractOver the past half‐century, cardiac pacing technology has adhered to one basic system. However, many pacemaker‐related complications are related to this system, particularly in terms of the power supply and leads. Here, for the first time a heart‐worn pacemaker strategy is presented, which allows batteryless powering and leadless pacing. The batteryless feature is attained via heart‐extrusion energy scavenging through a micromachined piezoelectric thick film transducer. A record in vivo output current of 30 µA is obtained in an adult swine by the implanted piezoelectric transducer, which can effectively drive a commercial cardiac pacemaker. The exocardial pacing method is demonstrated by this batteryless pacemaker functions without a device or leads placed within the cardiac chambers. Lead‐ or batteryless options are being explored as new basic features of cardiac pacemakers. The presented heart‐worn pacemaker strategy may be useful in future pacing technology.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.