Abstract
The method of measuring f(I) values is proposed as a measure of the adsorption capacity of adsorbents. The f(I) value, a characteristic figure for a particular system, can be obtained chromatogiaphically by measuring the progression rate of zone and the degree of packing of the column. For a system involving a unit concentration of trasns- p-methoxyazobenzene in benzene as standard solution and alumina as adsorbent, optimum conditions for the measurement of f(I) were established. The effects of day-light, pretreatment of column, temperature, flow rate of solution, amount of adsorbent and degree and mode of packing were studied Under optimum conditions, the value of f(I) was reproducible within ±2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.