Abstract

We present a ballistocardiogram (BCG) measurement system for obtaining the BCG signal. The performance of the BCG measurement system was tested, including the dynamic response frequency of the platform (32 Hz with a 75 kg load), and the signal to noise ratio of the system (minimum 23 dB). Factors that could influence BCG recording during measurement were investigated by collecting data from 48 healthy subjects (28 males and 20 females). The basic features of the BCG signal, such as the amplitude and time interval, were extracted, and differences by gender were analyzed. A correlation between the IJ amplitude (amplitude difference between I- and J-peak) and the respiration phase was found (R 2 = 0.49). The BCG signals for both sitting (SiBCG) and standing (StBCG) postures were obtained using this system. These showed that the StBCG amplitudes were larger than those of SiBCG but there was no significant difference in the time intervals. The robustness of our system was verified, and its potential for non-invasive cardiac activity monitoring was demonstrated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.