Abstract

To estimate geometric parameters in a skew-slit single photon emission computed tomography (SPECT) scan, a simple method is presented that uses a single point source and a backprojection process. The method employs one point source placed sufficiently far from the axis of rotation but in the field of view. A skew-slit geometry is considered where the estimated parameters include the focal lengths of both opaque planes and the location of multiple transaxial slits and a single axial slit. Computer simulations show that view-angle independent parameters can be uniquely estimated. Three experimental scans were performed, and the view-angle independent parameters were accurately estimated. Tilt, sagging, and mechanical shifts result in parameters that change depending on the view angle in the experimental scans. These view-angle dependent parameters were estimated using the single point-source technique, but reconstructed phantom images showed some error in the estimation of the view-angle dependent parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.