Abstract

We demonstrate a high-brightness compact 9 keV electron-impact microfocus x-ray source based on a liquid-gallium-jet anode. A approximately 30 W, 50 kV electron gun is focused onto the approximately 20 ms, 30 mum diameter liquid-gallium-jet anode to produce an approximately 10 microm full width at half maximum x-ray spot. The peak spectral brightness is >2 x 10(10) photons(s mm(2) mrad(2)x0.1% BW). Calculation and experiments show potential for increasing this brightness by approximately three orders of magnitude, making the source suitable for laboratory-scale x-ray crystallography and hard x-ray microscopy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.