Abstract
Flash x-ray (FXR) systems are used for dynamic radiography. Depending on the speed of the object, these systems typically require a very short pulse duration (∼25ns) for image acquisition without motion blur. The conventional Marx generators with zigzag discharge paths result in higher inductance; hence, they do not meet the requirement of shorter pulse duration (30-40ns) and low impedance (40-60 Ω) simultaneously. A coaxial Marx generator has been designed and developed, which is capable of generating 500kV peak voltages and 10kA peak current within a 40ns pulse duration. The CST simulation of the coaxial Marx generator has been carried out to validate the design parameters. The FXR electron beam diode is powered by this Marx generator. Experiments were carried out to measure the x-ray parameters like pulse width, source size, x-ray energy spectrum, penetration depth, and cone angle. The maximum measured x-ray dose was 62 mR at 1m distance from the source window. The x-ray radiograph demonstrates a penetration depth of 32mm in steel kept at 2.5m distance from the source for 500kV diode voltages.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.