Abstract

Standard flow cytometry protocols for CD34+ cell enumeration designed for fresh samples are not appropriate for cryopreserved products. Special protocols have been developed to remove the cryoprotectant by quickly washing a freshly thawed sample. Exposing cells to a large volume of hypotonic solution and subsequent washing process was hypothesized to cause lab-induced cell death. Moreover, standard gating strategies must be altered to avoid reporting falsely high viabilities. We developed a novel method whereby thawed samples were diluted step-wise to 1:2 by 3 additions of 1/3 sample volume using 1% Human Albumin in Dextran 40 (10% Low Molecular Weight Dextran in 0.9% NaCl) separated by 5 min between each addition. An additional 1:10 dilution was required to obtain a desired cell concentration for flow cytometry testing resulting in a 1:20 dilution. Twenty samples were tested simultaneously in a method comparison; the new method demonstrated significant increases in mean cell viabilities for white blood cells, hematopoietic progenitor cells, and T cells as well as reduced standard deviations for each parameter. Slow, step-wise dilutions of freshly thawed samples of cryopreserved apheresis products to 1:20 yielded higher and more precise viability measurements compared to quickly washing samples to remove DMSO.