A dilution refrigerator with two mixing chambers is described. When compared with a dilution refrigerator with only one mixing chamber the temperature range is extended to lower temperatures. The temperature reduction factor can be up to a factor 2.8. The results of an experiment are reported in which a magnetic temperature of 5.5 mK was obtained in the continuous mode and 3.5 mK in the single cycle mode, using a dilution refrigerator that reaches 13 mK in the continuous mode with one mixing chamber.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.