Abstract
Positron emission tomography (PET) is commonly used to detect tumours and assess the progress of cancer treatment by measuring standardized uptake values (SUV). Physiological motion during data acquisition can negatively impair obtained SUV. Simultaneous PET-MR combines metabolic PET information with high-resolution anatomical MR images and has the potential to minimize motion artefacts. A 3D MR-acquisition scheme is proposed which allows for the automatic detection and correction of non-rigid bulk motion shifts, i.e. changes of the patient’s position leading to non-rigid changes of the patient’s anatomy, in PET and MR images (both T1-weighted gradient echo and T2-weighted turbo spin echo images) without an increase in scan time. Six respiratory gated T1- and T2-weighted 3D data sets with an isotropic resolution of 1.5mm were obtained using a radial phase encoding (RPE) acquisition. Volunteers are asked to move the abdomen during data acquisition resulting in overall 19 movements at arbitrary time points. RPE allows for a retrospective reconstruction of dynamic 3D MR images with different temporal resolutions from the same data. Non-rigid bulk motion is detected and corrected from these dynamic images. A simultaneous PET acquisition is simulated to assess the effect of motion correction on image quality and SUV for lesions with different diameters. Every bulk motion shift was accurately detected and corrected. Non-rigid motion fields describing the different motion states were obtained with an accuracy of 1.71 ± 0.29mm. The PET simulation showed SUV errors of up to 67% due to bulk motion which could be reduced to 10% with the proposed motion compensated approach. The presented MR acquisition scheme yields both high resolution 3D anatomical data and highly accurate non-rigid motion information without an increase in scan time. This method strongly improves both MR and PET image quality and ensures an accurate assessment of tracer uptake.
Highlights
Positron emission tomography (PET) is commonly used to detect tumours and assess the progress of cancer treatment by measuring standardized uptake values (SUV)
A 3D MR-acquisition scheme is proposed which allows for the automatic detection and correction of non-rigid bulk motion shifts, i.e. changes of the patient’s position leading to non-rigid changes of the patient’s anatomy, in PET and MR images without an increase in scan time
Volunteers are asked to move the abdomen during data acquisition resulting in overall 19 movements at arbitrary time points
Summary
Positron emission tomography (PET) is commonly used to detect tumours and assess the progress of cancer treatment by measuring standardized uptake values (SUV). A 3D MR-acquisition scheme for non-rigid bulk motion correction in simultaneous PET-MR Christoph Kolbitsch1*, Claudia Prieto1, Charalampos Tsoumpas1,2, Tobias Schaeffter1 From PSMR14: 3rd Conference in PET/MR and SPECT/MR Kos Island, Greece.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.