Abstract

This study presents a simple process to assemble the 3D glass microprobe array. Glass microprobe with embedded silicon (ES) was fabricated by glass reflow process. In the 1st assembly, the chips with 2D glass microprobe array were bonded by parylene-C to form the 3D the microprobe array. In the 2nd assembly, the 3D probe array was then mounted on the silicon carrier. The ES is employed for alignment during the assembly, and also acts as the electrical routing for signal recording. In application, the impedance of this glass microprobe was measured. The impedance at 1kHz was 1.1MΩ. Action potential from crayfish nerve cord was also successfully recorded.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.