Abstract

BackgroundRapid Point-Of-Care Tests for Chlamydia trachomatis (CT) may reduce onward transmission and reproductive sexual health (RSH) sequelae by reducing turnaround times between diagnosis and treatment. The io® single module system (Atlas Genetics Ltd.) runs clinical samples through a nucleic acid amplification test (NAAT)-based CT cartridge, delivering results in 30min. MethodsProspective diagnostic accuracy study of the io® CT-assay in four UK Genito-Urinary Medicine (GUM)/RSH clinics on additional-to-routine self-collected vulvovaginal swabs. Samples were tested “fresh” within 10days of collection, or “frozen” at −80°C for later testing. Participant characteristics were collected to assess risk factors associated with CT infection. ResultsCT prevalence was 7.2% (51/709) overall. Sensitivity, specificity, positive and negative predictive values of the io® CT assay were, respectively, 96.1% (95% Confidence Interval (CI): 86.5–99.5), 97.7% (95%CI: 96.3–98.7), 76.6% (95%CI: 64.3–86.2) and 99.7% (95%CI: 98.9–100). The only risk factor associated with CT infection was being a sexual contact of an individual with CT. ConclusionsThe io® CT-assay is a 30-min, fully automated, high-performing NAAT currently CE-marked for CT diagnosis in women, making it a highly promising diagnostic to enable specific treatment, initiation of partner notification and appropriately intensive health promotion at the point of care.

Highlights

  • Rapid Point-Of-Care Tests for Chlamydia trachomatis (CT) may reduce onward transmission and reproductive sexual health (RSH) sequelae by reducing turnaround times between diagnosis and treatment

  • The objective of this study was to conduct a performance evaluation of the io® CT assay using genital samples collected from females attending UK Genito-Urinary Medicine (GUM)/Reproductive and Sexual Health (RSH) clinics

  • A total of 785 female participants were recruited from the different clinics (Fig. 1). 76 participants were excluded from the final analyses, conducted on 709 (90.3%) participants, for reasons including not fulfilling eligibility criteria, missing initial reference test data, and final invalid io® CT assay results

Read more

Summary

Introduction

Rapid Point-Of-Care Tests for Chlamydia trachomatis (CT) may reduce onward transmission and reproductive sexual health (RSH) sequelae by reducing turnaround times between diagnosis and treatment. The io® single module system (Atlas Genetics Ltd.) runs clinical samples through a nucleic acid amplification test (NAAT)-based CT cartridge, delivering results in 30 min. Methods: Prospective diagnostic accuracy study of the io® CT-assay in four UK Genito-Urinary Medicine (GUM)/ RSH clinics on additional-to-routine self-collected vulvovaginal swabs. Participant characteristics were collected to assess risk factors associated with CT infection. The only risk factor associated with CT infection was being a sexual contact of an individual with CT. Conclusions: The io® CT-assay is a 30-min, fully automated, high-performing NAAT currently CE-marked for CT diagnosis in women, making it a highly promising diagnostic to enable specific treatment, initiation of partner notification and appropriately intensive health promotion at the point of care

Objectives
Methods
Results
Conclusion