Abstract

For radiography applications using fast neutrons simultaneously with gammas we have developed a detector with 16 stilbene crystals in a 4×4 2D array with a 5 mm pitch and a depth of 25 mm. The crystal array is read out by Silicon photomultipliers and custom signal processing electronics. The detector prototype was tested using a custom D-D fast neutron generator at the Paul Scherrer Institute. By applying a pulse shape discrimination algorithm the detector is able to detect and distinguish fast neutrons and gammas simultaneously. Various attenuating samples placed between the source and detector with different materials and thicknesses were tested and the measured macroscopic fast neutron cross sections were compared to what was expected. Deviations were studied with the help of detailed Geant4 simulations. The detection efficiency for D-D fast neutrons was measured to be around 10%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.