Abstract

Introduction: The aim of present work is showing a procedure to evaluate the radiological monitor dose to pacemaker of patient undergoing treatment with CyberKnife® (Accuray, Sunnyvale, CA). Radiological monitor dose is dose coming from radiological images that are generated by x-ray tubes to monitor patient positioning during radiotherapy dose delivering. The generation of such images gives patient total dose a contribution that is not accounted by treatment planning system (tps); tps in fact just takes into account radiotherapy dose and not even radiological monitor dose. In some situations this cannot be overlooked and it has to be accounted for; one of these may be a patient with a pacemaker. For this device manufacturer likely indicates a threshold dose value over witch functioning is compromised. Our patient to be treated with CyberKnife® has a pacemaker with a dose threshold value indicated by manufacturer as 2 Gy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.