Abstract

A major design problem concerning high-frequency broad-band ultrasound imaging systems is caused by the strong dispersive attenuation of the tissue, which gives rise to images with inhomogeneous resolution and poor signal to noise ratio (SNR). To address the noise problem, strongly focused transducers with high energy density in a narrow focal region are utilized, which also provide more isotropic images due to improved lateral resolution. To account for the short depth of the focal area two suitable imaging conceptions are used: 1) synthetic aperture concept and 2) B/D-scan concept. To avoid the inhomogeneity of the images, different transmitter signals for each depth are applied, which are pseudoinversely prefiltered according to the transfer function of the tissue. To gain signal energy required for inverse filtering, a pulse compression technique with nonlinearly frequency modulated chirp signals is utilized. These procedures have been implemented in an ultrasound imaging system, which has been developed in the authors' laboratory for eye and skin examinations, It can be used with transducers in a frequency range from 20 to 250 MHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.