Abstract

Diagnosis and minimal residual disease (MRD) monitoring of chronic lymphocytic leukemia (CLL) by flow cytometry currently requires multiple antibody panels. We added CD23 and CD200 to the EuroFlowTM lymphoid screening tube (LST) to create a 10-color modified LST (mLST) capable of diagnosing typical CLL in a single tube. We then explored if the mLST could be used for MRD by comparing its performance to the European Research Initiative on CLL (ERIC) panel using spiked cryopreserved and fresh patient samples. Over 1 year of use in our clinical laboratory, the mLST diagnosed CLL without further immunophenotyping in 56% of samples with an abnormal clone. There was good agreement in MRD results between the mLST and ERIC panels. Therefore, the mLST can streamline CLL diagnosis by reducing technician time and the number of panels required. It may have the potential to screen for MRD in laboratories without access to dedicated panels (ERIC).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.