Abstract

This work presents a third-order high-pass sigma-delta modulator (HPSDM) for biomedical signal acquisition. The operational amplifier (op-amp) sharing and noise-coupling techniques are adopted to reduce the required quantity of op-amps and add a noise-shaping order, which can achieve low power consumption and high resolution. A novel switched-capacitor architecture is proposed to suppress the increasing in-band noise and alleviate the circuit sensitivity to capacitor mismatch in the high-pass integrator. The proposed HPSDM was fabricated in a 0.18-μm standard CMOS process. Measurement results reveal that the proposed HPSDM has a signal-to-noise and distortion ratio (SNDR) of 75.26/74 dB in 200 Hz bandwidth and consumes 1.52/0.8 μW under 1.2/1 V supply voltage, which can achieve a peak Schreier Figure-of-Merit of 156.45/157.98 dB and a peak Walden FoM of 0.802/0.488 pJ/conv.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.