Abstract

In this paper, an extensive work is done on the specific tune of a conventional ICP-MS for 99Tc atom counting. For this, two methods have been used and compared: the partial variable control method and the 5D Simplex method. Instrumental limits of detection of 0.2 and 0.8 ppt, respectively, were obtained. They are noticeably lower than that found with the automatic tune method of the spectrometer, 47 ppt, which shows the need of a specific tune when very low levels of 99Tc have to be determined. A study is presented on the mass interferences for 99Tc. Our experiments show that the formation of polyatomic atoms or refractory oxides as well as 98Mo hydrides seem to be irrelevant for 99Tc atom counting. The opposite occurs with the presence of isobaric interferences, i.e. 99Ru, and the effect of abundance sensitivity, or low-mass resolution, which can modify the response at m/ z=99 to a non-negligible extent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.