Abstract

were present in respectively 97%, 40%, and 60% using SE, and 92%, 33%, and 58% using the HD (p=NS). There was a good agreement between the 2 imaging devices for the sclerosis of all aortic valve cusps (right cusp 93.2%, k=0.89, left cusp 94.7%, k=0.88, non coronary cusp 81.4%, k=0.74). The agreement was also good for evaluation of MAC and ARS, respectively 89.7%, k=0.78 and 98.2%, k=0.96. During DSE wall motion abnormalities at rest and ischemic segments at peak were present respectively in 36 (50%) and 11 (23%) of patients. There was no correlation between the grade of AVS and the number of dysfunctional segments at rest (p= 0.20) or the number of ischemic segments at peak DSE (p=0.21). Conclusion: Hand-held echocardiography is a simple, reliable and inexpensive tool to screen patients for AVS, MAC and ARS. 927 The performance of hand-carried ultrasound devices for assessment of global and segmental left ventricular function is improved with the use of echocontrast.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.