Abstract

To verify the test-retest, inter and intraexaminer variability of the Compound Muscle Action Potential (CMAP) and Sensory Nerve Action Potential (SNAP) within our neurophysiology laboratory; to identify the technical variables that can affect the reproducibility of signals in the follow-up studies. Distal CMAP of Median and Ulnar nerves and SNAP of radial nerve (I ° finger), median nerve (I° and IV° fingers), ulnar nerve (IV and V fingers) of the right hand were recorded by 9 operators (5 technicians of neurophysiology and 4 physicians) in healthy volunteers. Each test was repeated 3 times randomly within 4 months. The latency, amplitude, and intensity of stimulus were collected. The montages were photographed. Our data showed a discrete stability of CMAP parameters mainly for ulnar nerve. On the contrary the amplitudes of SNAPs were fairly variables. The results suggest that it is crucial to check the neurophysiological techniques commonly applied in laboratories were several operators work. More time must be dedicated to a common training, to get consistent results useful in repeated trials.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.