Abstract

We report the first comparison of cardiovascular magnetic resonance imaging (CMR) at 1.5 T, 3 T and 7 T field strengths using steady state free precession (SSFP) and fast low angle shot (FLASH) cine sequences. Cardiac volumes and mass measurements were assessed for feasibility, reproducibility and validity at each given field strength using FLASH and SSFP sequences. Ten healthy volunteers underwent retrospectively electrocardiogram (ECG) gated CMR at 1.5 T, 3 T and 7 T using FLASH and SSFP sequences. B1 and B0 shimming and frequency scouts were used to optimise image quality. Cardiac volume and mass measurements were not significantly affected by field strength when using the same imaging sequence (P > 0.05 for all parameters at 1.5 T, 3 T and 7 T). SSFP imaging returned larger end diastolic and end systolic volumes and smaller left ventricular masses than FLASH imaging at 7 T, and at the lower field strengths (P < 0.05 for each parameter). However, univariate general linear model analysis with fixed effects for sequence and field strengths found an interaction between imaging sequence and field strength (P = 0.03), with a smaller difference in volumes and mass measurements between SSFP and FLASH imaging at 7 T than 1.5 T and 3 T. SSFP and FLASH cine imaging at 7 T is technically feasible and provides valid assessment of cardiac volumes and mass compared with CMR imaging at 1.5 T and 3 T field strengths.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.