Abstract
BackgroundTo evaluate feasibility and diagnostic performance of clinical 7-T magnetic resonance imaging (MRI) of the shoulder.MethodsEight patients with suspected lesions of the rotator cuff underwent 7-T MRI before arthroscopy. Image quality was scored for artifacts, B1+ inhomogeneities, and assessability of anatomical structures. A structured radiological report was compared to arthroscopy. In four patients, a visual comparison with pre-existing 1.5-T examinations was performed.ResultsRegarding image quality, the majority of the sequences reached values above the middle of each scoring scale. Fat-saturated proton density sequences showed least artifacts and best structure assessability. The most homogenous B1+ field was reached with gradient-echo sequences. Arthroscopy did not confirm tendinopathy/partial tear of supraspinatus in 5/8 patients, of subscapularis in 5/6, and of infraspinatus in one patient; only a partial lesion of the subscapularis tendon was missed. Pathologic findings of long bicipital tendon, acromioclavicular joint, glenohumeral cartilage, labrum, and subacromial subdeltoideal bursa were mainly confirmed; exceptions were one lesion of the long bicipital tendon, one subacromial bursitis, and one superior glenoid labrum anterior-to-posterior lesion, missed on 7-T MRI. Evaluating all structures together, sensitivity was 86%, and specificity 74%. A better contrast and higher image resolution was noted in comparison to previous 1.5-T examinations.Conclusions7-T MRI of the shoulder with diagnostic image quality is feasible. Overrating of tendon signal alterations was the main limitation. Although the diagnostic performance did not reach the current results of 3-T MRI, our study marks the way to implement clinical 7-T MRI of the shoulder.
Highlights
To evaluate feasibility and diagnostic performance of clinical 7-T magnetic resonance imaging (MRI) of the shoulder
Following the trend showed in studies comparing 1.5-T to 3-T MRI of the shoulder in the detection of partial- and full-thickness rotator cuff tears [8], we postulated that even more accuracy should be gained from further increased field strengths
Findings and diagnostic performance With 7-T MRI, we identified pathologic findings regarding the rotator cuff in every patient
Summary
To evaluate feasibility and diagnostic performance of clinical 7-T magnetic resonance imaging (MRI) of the shoulder. As a standard of care procedure, it is widely used at field strengths of 1.5 or 3 T and has proven high diagnostic accuracy in the diagnosis of rotator cuff tears [3]. Lazik-Palm et al European Radiology Experimental (2020) 4:10 are desired. These demands can either be solved by extending the examination time to a certain extent or by further increasing the magnetic field strength [7]. Following the trend showed in studies comparing 1.5-T to 3-T MRI of the shoulder in the detection of partial- and full-thickness rotator cuff tears [8], we postulated that even more accuracy should be gained from further increased field strengths
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.