Abstract
To assess the diagnostic performance of 64-section computed tomography angiography (CTA) for detection of cerebral aneurysms by comparison with the new gold standard three-dimensional (3D) digital subtraction angiography (DSA). A total of 128 patients who underwent both 64-section CTA and 3D DSA for suspected intracranial aneurysms were included. The location, size, and shape of the aneurysm were assessed and compared with 3D DSA results. The sensitivity, specificity, positive and negative predictive values of 64-section CTA were calculated by using 3D DSA as reference standard. According to the reference standard, a total of 145 aneurysms were detected at 3D DSA in 118 patients. Of these 145 aneurysms, 64-section CTA detected 139 aneurysms in 118 patients. The sensitivity of CTA for detection of aneurysm was 95.8% on a per-aneurysm basis. All missed aneurysms were non-causative very small aneurysms in patients harboring multiple aneurysms. On a per-patient basis, the sensitivity, specificity, positive and negative predictive values were 100%. Compared with the new gold standard 3D DSA, 64-section CTA offers high sensitivity and specificity for detection of intracranial aneurysms. It could be readily used as a screening imaging method for detection of intracranial aneurysms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.