Abstract
For modernization of the Russian national activity standard at the Ionizing Radiation Department of the D.I. Mendeleyev Institute for Metrology, a prototype 4π(LS)-γ(NaI) coincidence arrangement was created, and applied to the standardization of long-lived 60Co and 22Na and short-lived 56Mn and 18F. Efficiency variation was performed using “grey filters”, high voltage variation and variation of low threshold. The main metrological characteristics of the setup were determined: long-term stability, background, dead time, resolution time and temperature dependence. The results obtained have practical applications. The 18F solution with well-known activity is required for calibration of ionizing chamber used in nuclear medicine. The 56Mn is used for calibration of manganese bath equipment used in neutron laboratory. The results obtained are in good agreement with 4πγ(NaI)-counting and 4πβ(PC)-γ(NaI) methods of Russian national radioactivity standard. The combined uncertainty (k=2) of results was estimated in the range 1–2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.