Abstract

The electrocardiogram (ECG) is one of the most commonly measured biosignals. In particular, textile electrodes allow for the measuring of long-term ECG without skin irritation or other discomforts for the patient. Such textile electrodes, however, usually suffer from insufficient or unreliable skin contact. Thus, developing textile electrodes is impeded by the often-complicated differentiation between signal artifacts due to moving and breathing and artifacts related to unreliable skin contact. Here, we suggest a simple method of using 50/60 Hz power grid noise to evaluate the skin contact of different textile electrodes in comparison with commercial glued electrodes. We use this method to show the drying of wetted skin under an embroidered electrode as well as sweating of the originally dry skin under a coated electrode with high water vapor resistance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.