Abstract

Abstract Introduction There is a greater risk of delayed healing and revision surgery in atypical femoral fractures (AFFs). Although non-union in reamed intra-medullary (IM) nailing is relatively uncommon, it can lead to fracture of the nail and present a considerable challenge to the treating surgeon. Case Report We present a case of AFF treated with IM nail fixation. Metalwork failure of the nail prompted removal of distal locking screws and plating of the fracture site at 8 months. Failure of the plate-bone interface 3 months later led to further revision surgery to remove the broken metalwork. We focus on the significant operative difficulty encountered during removal of the failed IM metalwork and provide a novel practical solution to overcome this particular challenge - insertion of a conical cannulated screw remover under fluoroscopy guidance to create an interference fit with the distal segment of the broken nail. Discussion This case is an example of significant and unexpected intra-operative technical difficulty, requiring improvisation and teamwork to manage. We introduce an accessible and uncomplicated alternative to the existing techniques for removal of a broken femoral nail and in doing so hope to benefit peers and colleagues should they encounter similar difficulties in the future.