Abstract

In recent years the apprication field of digital subtraction angiography (DSA) is expanding. Following are the results of some of the technical problems discussed in connection with the clinical utilization of DSA. 1. After installation of DSA system, the measured brightness of C-arm type Toshiba image intensifier was Gx=0.36. 2. For removal purpose, the scattered radiation air gap method was used instead of grid in which a 15 cm distance was adequate between the patient and the image intensifier. As a consequence, the radiation exposure dose was 30% less than that of using a grid ratio 10 : 1,and the heat load on the X-ray tube was reduced as well. 3. Spatial and contrast resolution of DSA unit were measured using DSA phantom by an image matrix of 512×512. The former was limited to 1 lp/mm of lead bars and the latter, 1 mg/cm^2 of iodine density.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.