Abstract

We propose a new SFS (shape from shading) technique for improved 3D surface reconstruction and imaging of relatively smooth surface topography using the scanning electron microscope (SEM). The new arrangement of backscattered electrons detector plates allows decreasing the initial energy of the electron probe, which makes this SEM technique to be suitable for usage on radiation-sensitive samples like biological tissues. Experiments show high effectiveness of the method, which improves both the gradient sensitivity of the signal and the signal to noise ratio.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.