Abstract

New 3D-ring CZT systems with low- and medium-energy-range detectors allow for simultaneous dual-isotope lung scintigraphy. We compared 10-, 7-, 5-, and 3-minute acquisitions for 99mTc and 81mKr performed simultaneously on StarGuide CZT-SPECT/CT and reframed in 50 patients. Ventilation/perfusion mismatches were calculated (mean, 15.6% ± 28%), and Spearman correlation coefficients of mismatches were 0.994, 0.994, and 0.984 between 10- and 7-, 5-, and 3-minute acquisitions, respectively. No visual difference in image quality or final diagnosis was found. 3D-ring CZT-SPECT with low and medium energy range detectors allows ultrafast dual-isotope lung scintigraphy up to 3 minutes.