Abstract
The diagnosis of osteoporosis and the prevention of femur fractures is a major challenge for our society. However, the diagnosis performed in clinical routine from Dual Energy X-ray Absorptiometry (DXA) images is limited. This paper proposes a 3D reconstruction method of both the shape and the Bone Mineral Density (BMD) distribution of the proximal femur from routinely used DXA images. The reconstruction accuracy that can be obtained from single-view and multi-view DXA devices was assessed. This evaluation, from 20 bone specimens and simulated DXA images, highlighted a mean shape accuracy of 1.3mm and a BMD accuracy of 4.4% from a single-view DXA image. A multi-view configuration with 2 views (frontal-sagittal) appeared as a good compromise (mean shape accuracy of 0.9mm and BMD accuracy of 3.2%). We are currently using this method for in vivo clinical studies in order to improve the diagnosis of osteoporosis and the prevention of femur fractures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.