Abstract

We investigated the role of metal artifact reduction by taking 3D print impressions using 3D data of Computed Tomography (CT) images based on the algorithm applied. We manufactured a phantom of a human mandible tooth made of gypsum and nickel alloy to measure the metal artifacts. CT images were obtained by changing the phantom tube voltage and tube current. The signal intensity of the image generated by the metal artifacts before and after the iterative metal artifact reduction algorithm (iMAR) was measured. A 3D printing process was performed after converting the images, before and after iMAR application, into STL files using InVesalius version 3.1.1 by selecting the conditions that minimized the effect of the artifact. Regarding metal artifacts, the Hounsfield unit (HU) value showed low as the tube voltage increased. The iMAR-applied images acquired under the same conditions showed a significantly lower HU. The artifacts, in the form of flashes, persisted in the 3D-printed product of the image not subjected to iMAR, but were largely removed in the 3D-printed product following iMAR application. In this study, the application of iMAR and data acquired using high tube voltage eliminated a significant portion of the metal artifacts, resulting in an impression shape that was consistent with the human body.